Humidity Thermal Shock Chambers HZ-2004A
Equipment summary:
Humidity Thermal Shock Chambers is applied to test the materials in heat-resistance, cold-resistance, dry-resistance, moisture-resistance. It is simple to operate and program easy to edit. It can show the setvalues and operative time.
Corresponding standard:
1.International Electrotechnical Commission Standard:
IEC68-2-03_test method Ca_Steady damp-heat
IEC68-2-01_test method A_cold
IEC68-2-02_test method B_dry heat
2. Military Standards:
MIL-STD-810F-507.4 humidity
MIL-STD-810F-501.4 high temperature
MIL-STD-810F-502.4 low temperature
MIL-STD883C test method 1004.2 temperature and humidity cycling test
MIL-STD810D test method 502.2
MIL-STD810 test method 507.2 procedure 3
3. Japanese Industrial Standard:
JIS C60068-2-3-1987 test method Ca: Steady damp-heat
JIS C60068-2-2-1995 test method B:dry heat
JIS C60068-2-1-1995 test method A:low temperature
4. USA Semiconductor Industry Standard:
JESD22-A101-B-2004 Constant temperature and humidity test
JESD22-A103-C-2004 High Temperature Storage Test
JESD22-A119-2004 Low Temperature Storage Test
5. Chinese National Standard:
GB11158 Technical conditions for high-temperature testing box
GB10589-89 Technical conditions for low-temperature testing box
GB10592-89 Technical conditions for high and low temperature testing box
GB/T10586-89 technical conditions for humidity testing chamber
GB/T2423.1-2001 Low-temperature testing methods
GB/T2423.2-2001 High-temperature testing methods
GB/T2423.3-93 Testing Method for humidity Test Chamber
GB/T2423.4-93 Alternating hot and humid testing method
GB/T2423.22-2001temperature testing method
6. China’s National Military Environment Test Equipment
GJB150.3 high-temperature testing
GJB150.4 low-temperature testing
GJB150.9 hot and humid condition testing
Applicable industries:
Humidity Thermal Shock Chambers applied to control the product quality, such as electronic, plastic products,electricalappliances, instruments, food, vehicles, metals, chemicals, building materials, aerospace, medical care and so on.
Detailed parameters:
model |
HZ-80 (A~G) |
HZ-150 (A~G) |
HZ-225 (A~G) |
HZ-408 (A~G) |
HZ-608 (A~G) |
HZ -800 (A~G) |
HZ-1000 (A~G) |
Inside size WxHxD (cm) |
40×50*40 |
50x60x50 |
50x75x60 |
60x85x80 |
80x95x80 |
100x100x80 |
100x100x100 |
Outside size W*H*D (cm) |
90x136x94 |
100x146x104 |
100x161x117 |
110x171x137 |
130x181x137 |
150x186x137 |
150x186x157 |
Temperature range |
-70 degree ~ +100(150) degree |
||||||
Humidity range |
20%~98%R.H.(10%-98%R.H / 5%~98%R.H is the specific condition ) |
||||||
Accuracy/ uniformly |
±0.1 degree ±0.1%R.H./ ±1.0 degree ±3.0%R.H. |
||||||
Accuracy / volatility |
±1.0 degree ±2.0%R.H. / ±0.5 degree ±2.0%R.H. |
||||||
Heating / cooling time |
About 4.0/min(heat),about 1.0/min(cool) ( 5~10/mm is the specific condition ) |
||||||
Inside material |
Stainless steel SUS 304 # |
||||||
Insulation material |
High-temperature high-density amino acid ethyl ester foam insulation materials, |
||||||
Cooling system |
Air-cooled/ single-stage compressor(-20 degree) , wind, water cooled/two-stage compressor(-40~-70 degree) |
||||||
protection |
compressor overload switch, high and low pressure protection switch, over-humidity-temperature protection switches, fuses, warning system |
||||||
accessories |
recorder(purchase), view window, 50mm testing hole, PL lamp, clapboard, Dry and wet gauze ball |
||||||
controller |
TEMI300 digital controller from South Korea | ||||||
compressor |
France Taikang |
||||||
power |
1Φ 220VAC±10% 50/60Hz & 3Φ 380VAC±10% 50/60Hz |
Feature :
1.High-texture appearance, the surface handled through the matte stripes, and use of planar non-reactive handles, easy to operate, safe and reliable.
2.Rectangular complex glass window can be viewed the testing sample, windows with anti-sweat heater device that prevents water vapor condensation, and high-brightness PL light.
3.Double-insulated airtight container, which can effectively isolate the internal temperature of leakage.
4.With an external water supply system can be easy to supply water and automatically reuse.
5.Compressor circulatory system from French “Taikang”, more effective removal of oil between the condenser and the capillary and use environmentally refrigerants ( R23, R404, R507).
6.Controller imported LCD display screen can be displayed measuring values and setting the value and time.
7.Controller with multi-segment program edited, temperature and humidity can be controlled quickly (OUICK) or slope (SLOP) .
8. Built-in mobile pulley easy to move and has a strong positioning screw fixed position
Equipment Configuration:
Name | Material | Model | Area | Description |
box | Outside box | Stainless steel SUS304 | import | antirust |
Inside box | Stainless steel SUS304 | import | antirust | |
insulation | Rigid polyurethane foam,
ultra-fine glass fiber |
Hongkong | insulationdoor | |
Stainless steel SUS304 | import | Good sealing performance | ||
View window | 38026 insulating glass | Observed the testing sample | ||
Sample holder | Stainless steel ring | import | Diameter 8mm | |
Testing hole | Diameter 60MM | Wire through directly | ||
Heating system | Heating tube | Stainless steel heat sink | Hongkong | Storage’s temperature rise up |
Air circulation | fan | YYF100-25-4S | Japan | Storage back to ring |
Cooling system | compressor | Taikang | France | cooling |
condensator
|
CAJ2464 | France | Storage back to ring | |
coolant | R23/404/environmentally | United States | cooling | |
Electromagnetic valve | 1028/3 Castor | Italy | Refrigerant opening and closing | |
Expansion valve | TY2 Danfoss | United States | Refrigerator throttling | |
filters | 0835 Danfoss | United States | Filter oxide cooling pipe | |
Humidify system | Humidity tube | Stainless steel | 304 import | The effective protection of
laboratory humidity |
Humidity box | Made by oneself | Provide temperature |